ยังไม่มีสื่อแสดงผล
คุณสมบัติและข้อมูลเชิงเทคนิค
- Centrifugal, direct-drive blowers designed for maximum energy efficiency and minimal maintenance
- Sentinel™ Gold Microprocessor Control System supervises all functions and monitors airflow and pressures in real-time
- Type D2 Pass Chamber as per ISO 14644-7 with interlocked doors, adjustable purged duration, and time-delayed ingress/egress control allowing sufficient time for surface decontamination to minimize transfer of contamination
- Removable sliding tray (option for perforated or non-perforated) prevents operator fatigue during transfer procedures
- Electromagnetic Interlocking Doors with time delay effect ensures safety and containment between the Pass Chamber and the Process Chamber
- Return Filter (option to upgrade to safe change BIBO) below work zone filters the contaminated air immediately to minimize possibility of airborne contamination
- Foot switch provides hands-free access to opening of the Pass Chamber inner door minimizing operator fatigue during transfer procedures
- HEPA H14 filters with a typical efficiency of >99.995% at 0.1-0.2 microns provide superior ISO Class 5 air cleanliness
- LED Lamps deliver >650 as per NSF49 to the work surface for superior overall illumination
- With ISO Class 5 environment, the Process Chamber facilitates aseptic processing in an EU GMP Grade A condition
- Circular glove ports (200 mm diameter) with minimal crevices, no exposed bolts and nuts
- Frameless tempered glass front visor panel with highly rounded edges to minimize crevices and maximize door vision panel
- Single or upgradable to multiple, removable work trays for easy surface cleaning and decontamination
- Optional adjustable hydraulic stand, with casters wheels, elevates to accommodate user preference for sitting or standing work surface height
|
GENERAL SPECIFICATIONS Isoclean® Healthcare Platform Isolator (HPI-G3) (with filter below)
|
HPI-2G |
HPI-3G |
HPI-4G |
|
|
Process Chamber Nominal Size (Width) |
1215 mm (47.83") |
1520 mm (59.84") |
1825 mm (71.85") |
|
|
External Dimensions (Wx D x H) - with 2 pass chamber |
Without stand |
2420
x 845 x 1560 mm |
2725 x 845 x 1560 mm (107.28" x 33.27" x 61.42") |
3030
x 845 x 1560 mm |
|
With SPC-A/SPL-A (713mm, 28") |
2420
x 845 x 2273mm |
2725
x 845 x 2273 mm |
3030
x 845 x 2273 mm |
|
|
With SPC-B/SPL-A (860mm, 34") |
2420
x 845 x 2420mm |
2725
x 845 x 2420 mm |
3030
x 845 x 2420 mm |
|
|
With STL/STC (660 to 885 mm) |
2420
x 845 x 2220 to 2445 mm |
2725
x 845 x 2220 to 2445 mm |
3030
x 845 x 2220 to 2445 mm |
|
|
With SHM (685 to 935 mm) |
2420
x 845 x 2245 to 2495 mm |
2725
x 845 x 2245 to 2495 mm |
3030
x 845 x 2245 to 2495 mm |
|
|
Process Chamber Internal Dimension (W x D x H) |
1215
x 624 x 610 mm |
1520
x 624 x 610 mm |
1825
x 624 x 610 mm |
|
|
Pass Chamber Internal Dimension (W x D x H) |
577 x 410 x 320mm 22.72" x 16.14" x 12.60") |
|||
|
Pass Chamber Door Dimension (W x H) |
Inner door |
428 x 400 mm (16.85" x 15.75") |
||
|
Outer door |
475 x 350 mm (18.70" x 13.78") |
|||
|
Airflow Regime |
Factory Configured Recirculating or Single Pass |
|||
|
Pressurization |
Factory Configured Positive or Negative Pressure |
|||
|
Glove Port Diameter |
200
mm (Circular) |
|||
|
Glove Port Quantity |
2 |
3 |
4 |
|
|
Chamber Environment |
ISO Class 5 for all chambers (Grade A) |
|||
|
Process Chamber Downflow Velocity |
0.4 +/- 20% m/s (1.31 fps) |
|||
|
Pre-filter |
G4, panel, polyester fiber media |
|||
|
Downflow and Exhaust Filter Type |
HEPA
(H14) Filter with Integral Metal Guards and Filter Frame Gaskets; Fully
Compliant With EN 1822 (H14) and |
|||
|
Filter Efficiency |
>99.995% for particle 0.1-0.2 microns (MPPS, as per EN1822) |
|||
|
Lighting Level |
>650 Lux |
|||
|
Sound Level |
≤ 67 dBA |
|||
|
Isolator Construction |
Main Body |
1.2 mm (0.05”) 18 gauge electro-galvanized steel with white oven-baked epoxy-polyester antimicrobial powder-coated finish |
||
|
Work Tray |
1.5 mm (0.06”) 16 gauge stainless steel, type 316L, with 4B finish |
|||
|
Electrical |
220-240V, AC, 50 Hz, 1Ø |
HPI-2G8-_ _ _ - _ |
HPI-3G8_-_ _ _ - _ |
HPI-4G8_-_ _ _ - _ |
|
Cabinet Full Load Amps (FLA) |
10 A |
12.1 A |
12.1 A |
|
|
Optional Outlets FLA |
10 A (5A per outlet) |
10 A (5A per outlet) |
10 A (5A per outlet) |
|
|
Cabinet Nominal Power |
493 W |
TBD |
TBD |
|
|
Cabinet BTU |
1682 BTU/hr |
TBD |
TBD |
|
|
110-120V, AC, 50 Hz, 1Ø |
HPI-2G9-_ _ _ - _ |
HPI-3G9_-_ _ _ - _ |
HPI-4G9_-_ _ _ - _ |
|
|
Cabinet Full Load Amps (FLA) |
20 A |
20 A |
20 A |
|
|
Optional Outlets FLA |
10 A (5A per outlet) |
10 A (5A per outlet) |
10 A (5A per outlet) |
|
|
Cabinet Nominal Power |
TBD |
TBD |
TBD |
|
|
Cabinet BTU |
TBD |
TBD |
TBD |
|
|
Options/ Accessories |
Adjustable Hydraulic Stand |
✓ |
✓ |
✓ |
|
Carbon Filter |
✓ |
✓ |
✓ |
|
|
CCTV |
✓ |
✓ |
✓ |
|
|
Drain |
✓ |
✓ |
✓ |
|
|
Electrical Outlet |
✓ |
✓ |
✓ |
|
|
Glove Leak Tester |
✓ |
✓ |
✓ |
|
|
IV Bar with S hooks |
✓ |
✓ |
✓ |
|
|
UV Lamp |
✓ |
✓ |
✓ |
|
|
Rear View Screen Adaptation |
✓ |
✓ |
✓ |
|
|
Single-piece Trays |
✓ |
✓ |
✓ |
|
|
Multiple-piece Trays |
✓ |
✓ |
✓ |
|
|
Anti-blow Back Valve (ABBV) |
✓ |
✓ |
✓ |
|
|
Minibioatom decontamination |
x |
x |
x |
|
|
Automated Pressure Hold Test |
x |
x |
x |
|
|
Sharps Disposal |
✓ |
✓ |
✓ |
|
|
Shipping Weight** |
835 kg (1840.47 lbs) |
967 kg (2133.33 lbs) |
1100 kg (2425.08 lbs) |
|
|
Shipping Dimensions, Maximum (W x D x H) |
2600
x 1050 x 2512 mm |
TBD |
3200
x 1050 x 2512 mm |
|
Optimized Solution for Sterile/Aseptic Applications
The Isoclean® Healthcare Platform Isolator (HPI-G3) facilitates the isolation of a product or process while providing the required sterile/aseptic environment. It is factory-configured to operate at positive or negative pressure in single or recirculating airflow. This equipment provides a comprehensive range of personnel and product protection in addition to the surrounding work areas and the environment.